Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf

Published May 13, 2020 · Recall initiated: April 7, 2020

Summary

FDA records medical device Class II recall Z-1888-2020, published May 13, 2020 and initiated April 7, 2020. Recalling firm: Inpeco S.A.; Status: Ongoing; Reported quantity: 23 units; Lot: Serial Numbers: ACP.357, ACP.388, ACP.396, ACP.397, ACP.399, ACP.400, ACP.402, ACP.404, ACP.405, ACP.407, ACP.408, ACP.409, ACP.410, ACP.412, ACP.413, ACP.416, ACP.417, ACP.418, ACP.419, ACP.420, ACP.421, ACP.422, ACP.423. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Inpeco S.A.
Distributor
Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri
Lot
Serial Numbers: ACP.357, ACP.388, ACP.396, ACP.397, ACP.399, ACP.400, ACP.402, ACP.404, ACP.405, ACP.407, ACP.408, ACP.409, ACP.410, ACP.412, ACP.413, ACP.416, ACP.417, ACP.418, ACP.419, ACP.420, ACP.421, ACP.422, ACP.423
Risk classification
Class II
Area (FDA)
device
Country
Switzerland

Related recalls

Questions about this recall

Was Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf recalled?

Yes. FDA published recall Z-1888-2020 on May 13, 2020. Reported hazard: The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.

What should I do if I have this product?

Voluntary: Firm initiated · Letter