AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Non-Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.

Published July 11, 2012 · Recall initiated: May 29, 2012

Summary

FDA records medical device Class I recall Z-1900-2012, published July 11, 2012 and initiated May 29, 2012. Recalling firm: Carefusion 2200 Inc; Status: Terminated; Reported quantity: N/A; Lot: product codes 0260-503; 0778-503, 10392-503, 10416-503, 10448-507, 10734-503, 10852-503, 1209-503, 1535-503, 1853-503, 5599-503, 5643-503, 5898-503, 6189-503, 6800-503 and RC41-00097; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Carefusion 2200 Inc
Distributor
Worldwide distribution: USA (nationwide) including Puerto Rico and internationally to: Canada, Egypt, France, Germany, Great Britain, Italy, Kuwait, Mexico, Portugal, Saudi Arabia, Spain, Taiwan and United Arab Emirates.
Lot
product codes 0260-503; 0778-503, 10392-503, 10416-503, 10448-507, 10734-503, 10852-503, 1209-503, 1535-503, 1853-503, 5599-503, 5643-503, 5898-503, 6189-503, 6800-503 and RC41-00097; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Non-Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient. recalled?

Yes. FDA published recall Z-1900-2012 on July 11, 2012. Reported hazard: The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter