Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter
Published May 13, 2020 · Recall initiated: July 8, 2019
Summary
FDA records medical device Class II recall Z-1905-2020, published May 13, 2020 and initiated July 8, 2019. Recalling firm: Medtronic Navigation, Inc.; Status: Terminated; Reported quantity: 4654 units; Lot: All kits with lot numbers from 0211041602 through 0217695790. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Navigation, Inc.
- Distributor
- US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK
- Lot
- All kits with lot numbers from 0211041602 through 0217695790
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801075, Sterile EO, 5/Tray 12 Pk., RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., RxonlyOctober 28, 2026
- FDATibial Tray Nonporous A Size 2; Model # PN 162-1200-A; Material: Ti6Al4V.May 13, 2020
- FDAGentell Honey Hydrogel 2"x2" Saturated Gauze- A hydrating wound dressing containing hydrogel and honey (100%May 13, 2020
Questions about this recall
Was Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter recalled?
Yes. FDA published recall Z-1905-2020 on May 13, 2020. Reported hazard: The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.
What should I do if I have this product?
Voluntary: Firm initiated · Letter