FDAFDA · Food and drugs·Recall number Z-1912-2024

Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1. Software:TKA 2.0.fr.1+TKA 1.0.1.fr.1+PKA3.0.2.fr. Application Part Number: 700002190743-04

Published June 5, 2024 · Recall initiated: April 15, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako System restart is not performed prior to switching between applications (i.e. TKA to THA), resulting in delay in treatment

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Orthopaedics
Distributor
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, Thailand, Turkey, UAE, and UK
Lot
UDI-DI: (01)07613327629095(10). Software:TKA 2.0.fr.1+TKA 1.0.1.fr.1+PKA3.0.2.fr. Application Part Number: 700002190743-04. Revision: AA & AB.
Risk classification
Class II
Area (FDA)
device
Country
United States