Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region.
Published July 9, 2014 · Recall initiated: June 9, 2014
Summary
FDA records medical device Class I recall Z-1920-2014, published July 9, 2014 and initiated June 9, 2014. Recalling firm: Medtronic Neurosurgery; Status: Terminated; Reported quantity: 3,026 units; Lot: Lot No. 206923217 207096627 207096630 207187691 207224897 207269986 207466017 207565424 207565425 207659575 207659576 207739874 207875274 207945037 207982847. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Neurosurgery
- Distributor
- Worldwide Distribution -- USA, Australia, Belgium, Canada, Croatia, Cyprus, Denmark, Egypt, Finland, France, Greece, Iran, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Switzerland, Turkey, and United Kingdom.
- Lot
- Lot No. 206923217 207096627 207096630 207187691 207224897 207269986 207466017 207565424 207565425 207659575 207659576 207739874 207875274 207945037 207982847
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVentriculostomy Kit, REF: 46154July 29, 2026
- FDAExacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 EXACTADecember 25, 2024
- FDABecker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKERDecember 25, 2024
- FDAMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Lumbar Catheter, REF 46917,March 6, 2024
- FDAThe Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationaryJuly 9, 2014
- FDAMedtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Catalog No. 46914. IntendedJuly 9, 2014
Questions about this recall
Was Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region. recalled?
Yes. FDA published recall Z-1920-2014 on July 9, 2014. Reported hazard: Medtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
What should I do if I have this product?
Voluntary: Firm initiated · Letter