Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device
Published June 19, 2024 · Recall initiated: May 8, 2024
Summary
FDA records medical device Class I recall Z-1921-2024, published June 19, 2024 and initiated May 8, 2024. Recalling firm: Thoratec Corp.; Status: Ongoing; Reported quantity: 4,842; Lot: UDI-DI: 00813024010142, 00813024010852. All Serial Numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Thoratec Corp.
- Distributor
- Worldwide distribution. US nationwide including Puerto Rico, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, BULGARIA, SWITZERLAND, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, ESTONIA, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GUADELOUPE, GREECE, IRELAND, ISRAEL, IRAN, ISLAMIC REPUBLIC OF, ITALY, KUWAIT, LEBANON, LITHUANIA, LUXEMBOURG, MACEDONIA, MARTINIQUE, NETHERLANDS, NORWAY, OMAN, POLAND, PALESTINE, PORTUGAL, QATAR, ROMANIA, SERBIA, RUSSIAN FEDERATION, SWEDEN, SLOVENIA, TUNISIA, TURKEY, SOUTH AFRICA, CANADA, AUSTRALIA, HONG KONG, MALAYSIA, SINGAPORE, SOUTH KOREA, TAIWAN, THAILAND, JAPAN, BANGLADESH, INDIA, PAKISTAN, ARGENTINA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, MEXICO, and BRAZIL
- Lot
- UDI-DI: 00813024010142, 00813024010852. All Serial Numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGrande Grande Primo Y45 Cultured Dairy Product. Product Code: 32400. Grade A. Pasteurized. Net Wt 50 lbs.June 19, 2024
- FDAPreview Strep A, Streptococcal A rapid test kits, Part Numbers: a) PRE-STA-25, b) PRE-STA-A-10June 19, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnlyMay 8, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnlyMay 8, 2024
- FDAHeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief,March 27, 2024
- FDAHeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REFMarch 27, 2024
Questions about this recall
Was Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device recalled?
Yes. FDA published recall Z-1921-2024 on June 19, 2024. Reported hazard: System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected
What should I do if I have this product?
Voluntary: Firm initiated · Letter