Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.
Published May 10, 2017 · Recall initiated: March 30, 2015
Summary
FDA records medical device Class II recall Z-1923-2017, published May 10, 2017 and initiated March 30, 2015. Recalling firm: Merge Healthcare, Inc.; Status: Terminated; Reported quantity: 23 sensors (PhaseIn Units); Lot: ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Merge Healthcare, Inc.
- Distributor
- US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.
- Lot
- ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMerge PACS and Merge OrthoPACS, medical imaging diagnostic workstationSeptember 2, 2026
- FDAMerge OrthoPACS, medical imaging diagnostic workstationAugust 12, 2026
- FDAMerge PACS, medical imaging diagnostic workstationAugust 12, 2026
- FDAVERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is anMarch 26, 2025
- FDAThe JETSTREAM is a rotational atherectomy catheter system designed for use in debulking and treating vascularMay 10, 2017
- FDASecond generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part NumberMay 10, 2017
Questions about this recall
Was Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system. recalled?
Yes. FDA published recall Z-1923-2017 on May 10, 2017. Reported hazard: The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone