Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.
Published August 14, 2013 · Recall initiated: July 1, 2013
Summary
FDA records medical device Class II recall Z-1924-2013, published August 14, 2013 and initiated July 1, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 5423; Lot: TACR Flex Reagent cartridge, material #10444938, (DF107) lot BB4087 with expiration date 3/28/14 and its linked calibrator, material #10445012, (DC 107) lot 3BD029 with expiration date 03/01/14. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- USA Nationwide Distribution including the states of: TX, NC, GA, TN, UT, NJ, MI, VA, AR, LA, MO, FL, AZ, CA, DC, PA, NH, WI, SD, SC, NM, MD, and IN.
- Lot
- TACR Flex Reagent cartridge, material #10444938, (DF107) lot BB4087 with expiration date 3/28/14 and its linked calibrator, material #10445012, (DC 107) lot 3BD029 with expiration date 03/01/14.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDABD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterileAugust 14, 2013
- FDADimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass Creatine Kinase MB isozyme CalibratorAugust 14, 2013
Questions about this recall
Was Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients. recalled?
Yes. FDA published recall Z-1924-2013 on August 14, 2013. Reported hazard: Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter