FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.
Published July 18, 2012 · Recall initiated: February 20, 2012
Summary
FDA records medical device Class I recall Z-1925-2012, published July 18, 2012 and initiated February 20, 2012. Recalling firm: Maquet Cardiovascular Us Sales, Llc; Status: Terminated; Reported quantity: 284 units; Lot: Model Number :6677300 Serial Number : 1170. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Distributor
- Worldwide distribution: USA (nationwide) state of GA only and countries of: Argentina, Australia, Austria, Belgium, China, Czech Republic, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, Spain, Sweden incl Maquet Critical Care, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
- Lot
- Model Number :6677300 Serial Number : 1170
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - ProductDecember 23, 2020
- FDAHLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 -December 23, 2020
- FDAYuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before,October 28, 2020
- FDAMaquet/Getinge Rotaflow System Console US (115 V / 60 Hz) - Product Usage: Instructions for Use - CentrifugalSeptember 23, 2020
- FDABanner Mountain Sprouts; Clover Sprouts: California Certified Organic Sprouts. 100% certified organic seedsJuly 18, 2012
- FDAAlere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB,July 18, 2012
Questions about this recall
Was FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. recalled?
Yes. FDA published recall Z-1925-2012 on July 18, 2012. Reported hazard: A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.
What should I do if I have this product?
Voluntary: Firm initiated · Letter