FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.

Published July 18, 2012 · Recall initiated: February 20, 2012

Summary

FDA records medical device Class I recall Z-1925-2012, published July 18, 2012 and initiated February 20, 2012. Recalling firm: Maquet Cardiovascular Us Sales, Llc; Status: Terminated; Reported quantity: 284 units; Lot: Model Number :6677300 Serial Number : 1170. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Maquet Cardiovascular Us Sales, Llc
Distributor
Worldwide distribution: USA (nationwide) state of GA only and countries of: Argentina, Australia, Austria, Belgium, China, Czech Republic, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, Spain, Sweden incl Maquet Critical Care, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
Lot
Model Number :6677300 Serial Number : 1170
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. recalled?

Yes. FDA published recall Z-1925-2012 on July 18, 2012. Reported hazard: A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.

What should I do if I have this product?

Voluntary: Firm initiated · Letter