HDL-Cholesterol plus gen.3.; and Small Supersized Intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of HDL-cholesterol concentration in serum and plasma.

Published July 8, 2015 · Recall initiated: May 7, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
US (nationwide) including PR.
Lot
Catalog Numbers: 05401488190; 0439980319; 05168805190; 04399803190, 04878205160.