The Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients.
Published May 13, 2020 · Recall initiated: February 14, 2020
Summary
FDA records medical device Class II recall Z-1927-2020, published May 13, 2020 and initiated February 14, 2020. Recalling firm: ICU Medical, Inc.; Status: Terminated; Reported quantity: 123 units; Lot: All serial numbers of Cogent HMS, List Number(s): 58400-000; 58400-000R; 58403-000; 58401-000 with Software Versions 1.3.3.42 and prior are impacted. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm identified software issues which leads to the patient ID and patient information not being stored in the system.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- US Nationwide distribution including in the states of AR, CA, FL, IL, MD, NC, NE, NY, TX, VT, WA and WI.
- UPC
- 00840619079499
- Lot
- All serial numbers of Cogent HMS, List Number(s): 58400-000; 58400-000R; 58403-000; 58401-000 with Software Versions 1.3.3.42 and prior are impacted.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAOncology sets with drip chambersSeptember 23, 2026
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAIV Non-Dedicated Gravity Administration Sets with drip chambersSeptember 23, 2026
- FDAMemantine Hydrochloride Extended-Release Capsules, 28 mg, 100 capsules per unit dose cartons, Rx only,May 20, 2020
- FDAPhilips Brilliance iCT SP Computed Tomography X-Ray System; Model # 728311May 13, 2020
Questions about this recall
Was The Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients. recalled?
Yes. FDA published recall Z-1927-2020 on May 13, 2020. Reported hazard: Firm identified software issues which leads to the patient ID and patient information not being stored in the system.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 00840619079499.