FDAFDA · Food and drugs·Recall number Z-1928-2014
Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems, Model No. 98202 The Alere Triage BNP test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications: ¿ as an aid in the diagnosis of congestive heart failure (also referred to as heart failure) ¿ as an aid in the assessment of severity of congestive heart failure ¿ for the risk stratification of patients with acute coronary syndromes ¿ for the risk stratification of patients with heart failure The Alere Triage BNP Calibrators are intended to calibrate the Alere Triage BNP test for the quantitative determination of BNP levels in human EDTA plasma using the Beckman Coulter Access Family of Immunoassay Systems.
Published July 2, 2014 · Recall initiated: May 30, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Alere San Diego is recalling Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems because the bottle labeled as Calibrator Level 5 actually contains Calibrator Level 4. This results in a calibration failure when attempting to calibrate using a calibrator kit which contains the affected Level 5 Calibrator.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Alere San Diego, Inc.
- Distributor
- Worldwide Distribution - Nationwide and the countries of Australia, Canada, China, EU, Mexico, Russia, Singapore, South Africa, South Korea, Taiwan and United Arab Emirates.
- Lot
- Lot #328163
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States