Triage D-dimer PN 98100 Used as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.
Published July 18, 2012 · Recall initiated: May 22, 2012
Summary
FDA records medical device Class I recall Z-1932-2012, published July 18, 2012 and initiated May 22, 2012. Recalling firm: Alere San Diego, Inc.; Status: Terminated; Reported quantity: 1,972 kits; Lot: Phase 1: W50069B, W50070B, W50071B, W50541B, W50551B, W50552B, and W50556B. Phase 2: W50550B. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alere San Diego, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Phase 1: W50069B, W50070B, W50071B, W50541B, W50551B, W50552B, and W50556B. Phase 2: W50550B
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine TestOctober 7, 2026
- FDAFisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + UrineOctober 7, 2026
- FDAMcKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine,October 7, 2026
- FDAFisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman CoulterOctober 7, 2026
- FDAAlere Triage BNP PN 98000XR Intended to be used as an aid in the diagnosis and assessment of severity ofJuly 18, 2012
- FDAMOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronicJuly 18, 2012
Questions about this recall
Was Triage D-dimer PN 98100 Used as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism. recalled?
Yes. FDA published recall Z-1932-2012 on July 18, 2012. Reported hazard: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter