FDAFDA · Food and drugs·Recall number Z-1942-2026

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Published May 13, 2026 · Recall initiated: March 25, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Ventec Life Systems, Inc.
Distributor
US: OH, GA, CA, MO
Lot
UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532
Risk classification
Class I
Area (FDA)
device
Country
United States