Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.
Published June 12, 2024 · Recall initiated: March 27, 2024
Summary
FDA records medical device Class II recall Z-1946-2024, published June 12, 2024 and initiated March 27, 2024. Recalling firm: Newport Corp; Status: Ongoing; Reported quantity: 19; Lot: P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696AF, 5691AF, 5692AF, 5693AF, 5694AF (9 of them in Irvine warehouse) P/N: N6-LPL-944-080 Power supply S/N: 23060081, 23060079. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
What to do
FDA Mandated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Newport Corp
- Distributor
- US Nationwide
- Lot
- P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696AF, 5691AF, 5692AF, 5693AF, 5694AF (9 of them in Irvine warehouse) P/N: N6-LPL-944-080 Power supply S/N: 23060081, 23060079
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment ofJune 12, 2024
- FDALeica HistoCore PEGASUS Tissue Processor, REF 14048858005.June 12, 2024
- FDAAlaris Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite Y Sites REF 10013072June 12, 2024
- FDAHA FlexTrak-Patient transport functionality to transport the patient from the preparation area to the MRJune 12, 2024
- FDAArial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and AppleJune 12, 2024
- FDAHA FlexTrak-Patient transport functionality to transport the patient from the preparation area to the MRJune 12, 2024
Questions about this recall
Was Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone. recalled?
Yes. FDA published recall Z-1946-2024 on June 12, 2024. Reported hazard: According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
What should I do if I have this product?
FDA Mandated · Letter