AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).
Published May 6, 2026 · Recall initiated: March 27, 2026
Summary
FDA records medical device Class I recall Z-1947-2026, published May 6, 2026 and initiated March 27, 2026. Recalling firm: North American Rescue LLC.; Status: Ongoing; Reported quantity: 21; Lot: Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- North American Rescue LLC.
- Distributor
- US: CA, WA
- Lot
- Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number:May 6, 2026
- FDAMedtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatmentMay 6, 2026
- FDAJUNCTIONAL EMERGENCY TREATMENT TOOL (JETT), REF: 30-0088, a component inside kits: KIT, TCCC Training-October 8, 2025
- FDARange Trauma Kit ORG - Product Code 80-0213, 80-0299, 80-0298, North American RescueDecember 14, 2016
- FDAUSCG Boat Response Kit - Product Code 80-0353, North American RescueDecember 14, 2016
- FDARange Trauma Kit Hardcase - Product Code 85-0889, North American RescueDecember 14, 2016
Questions about this recall
Was AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100). recalled?
Yes. FDA published recall Z-1947-2026 on May 6, 2026. Reported hazard: Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
What should I do if I have this product?
Voluntary: Firm initiated