AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and Accessories. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators. The device is model FC03XX series. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators.
Published July 18, 2012 · Recall initiated: June 15, 2012
Summary
FDA records medical device Class II recall Z-1951-2012, published July 18, 2012 and initiated June 15, 2012. Recalling firm: Encision, Inc.; Status: Terminated; Reported quantity: 48 boxes; Lot: RDB, RJB, RJE, RJH, RLG, RLI, RLJ, RLK, RLO, RLP, RLR, RLQ, SBF, SBJ. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Encision is performing a field correction to their AEM Disposable Electrodes due to two consumer complaints reporting unanticipated alarms from the AEM monitor while using the device. This issue could result delays or complications during surgical procedures.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Encision, Inc.
- Distributor
- Worldwide Distribution - USA including CA, TX, GA, NY, NC, OH), and New Zealand.
- Lot
- RDB, RJB, RJE, RJH, RLG, RLI, RLJ, RLK, RLO, RLP, RLR, RLQ, SBF, SBJ.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADrager PT 4000 Phototherapy SystemJuly 18, 2012
- FDACIVCO VirtuTRAX Instrument Navigator, Reusable non-sterile general purpose driveBAY, trakSTAR, electromagneticJuly 18, 2012
- FDASara Plus Mobile Patient Lifter; manufactured for Arjo by Medibo Medical Products N.V., Heikant 5, B-3930July 18, 2012
- FDACIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3DJuly 18, 2012
- FDAChorus Mobile Patient Lifter; manufactured by Arjo Ltd., St. Catherine Street, Gloucester, GL1 2SL, England,July 18, 2012
- FDACIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3DJuly 18, 2012
Questions about this recall
Was AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and Accessories. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators. The device is model FC03XX series. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators. recalled?
Yes. FDA published recall Z-1951-2012 on July 18, 2012. Reported hazard: Encision is performing a field correction to their AEM Disposable Electrodes due to two consumer complaints reporting unanticipated alarms from the AEM monitor while using the device. This issue could result delays or complications during surgical procedures.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit