ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581), SMN 10378677. Used on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP platforms. For in vitro diagnostic use in the quantitative determination of total homocysteine (HCY) in serum or plasma.
Published July 9, 2014 · Recall initiated: June 25, 2013
Summary
FDA records medical device Class II recall Z-1960-2014, published July 9, 2014 and initiated June 25, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc; Status: Terminated; Reported quantity: 51 units; Lot: Lot #/Exp. Date: 88D1524 01 Jun 2014 88D3303 26 Nov 2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
ADVIA Centaur Homocysteine dilution recovery on the diluent identified the percent recovery for the 1:10 patient sample dilutions lower than reported in the Instructions for Use (IFU).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of : Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Italy Kazakhstan, Latvia, Liechtenstein, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland ,Turkey, Unit.Arab Emir and Vatican
- Lot
- Lot #/Exp. Date: 88D1524 01 Jun 2014 88D3303 26 Nov 2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:10310308January 29, 2020
- FDASiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 10492399May 8, 2019
- FDAAtellica IM aTG (500 test kit) Siemens Material Number (SMN): 10995462May 8, 2019
- FDASiemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461May 8, 2019
- FDAAtrium Express Dry Seal Chest Drain ATS Blood Recovery, Catalog Number: 4050-100N. Autotransfusion Apparatus.July 9, 2014
- FDAMIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm);July 9, 2014
Questions about this recall
Was ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581), SMN 10378677. Used on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP platforms. For in vitro diagnostic use in the quantitative determination of total homocysteine (HCY) in serum or plasma. recalled?
Yes. FDA published recall Z-1960-2014 on July 9, 2014. Reported hazard: ADVIA Centaur Homocysteine dilution recovery on the diluent identified the percent recovery for the 1:10 patient sample dilutions lower than reported in the Instructions for Use (IFU).
What should I do if I have this product?
Voluntary: Firm initiated · Letter