Terumo® Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo® Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.

Published August 21, 2013 · Recall initiated: July 26, 2013

Summary

FDA records medical device Class II recall Z-1968-2013, published August 21, 2013 and initiated July 26, 2013. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 1657; Lot: Catalog number: 801763. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide Distribution - USA (nationwide) including the countries of Australia, Dubai, Hong Kong, Indonesia, Singapore, China, Mexico, British Columbia, Taiwan, Thailand, Chile, Malaysia, Japan, Belgium, United Arab Emirates (UAE), South Korea, Manila and Canada.
Lot
Catalog number: 801763
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Terumo® Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo® Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment. recalled?

Yes. FDA published recall Z-1968-2013 on August 21, 2013. Reported hazard: The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.

What should I do if I have this product?

Voluntary: Firm initiated · Letter