220/240V AC, Advanced Perfusion System 1 Product Usage: The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns TM or similar equipment.
Published August 21, 2013 · Recall initiated: July 26, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
TCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm condition occurs that displays a "999" pressure reading and results in stopping a pump, this pressure display can be cleared and the pump re-started by disconnecting the pressure transducer from the system.
What to do
Voluntary: Firm initiated
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Worldwide Distribution - US Nationwide in the states of :AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MSMT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and countries of AUSTRALIA, UNITED ARAB EMIRATES (UAE),Hong Kong,Indonesia, Singapore, Taiwan, Thailand,Philippines, CHILE,Malaysia, South Korea, BELGIUM,Japan,Mexico,CANADA
- Lot
- Catalog number: 801764
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States