Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);

Published July 18, 2012 · Recall initiated: June 5, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
Lot
Part numbers 3573-1 (angle handle) all lots including 64PY, 67YA, 69FF, 71GB, 72JL, 73GY
Risk classification
Class II
Area (FDA)
device
Country
United States