Langston Dual Lumen Catheters, 6Fr, REF 5550 and 5540. The Vascular Solutions Langston Dual Lumen Pressure Monitoring catheter is used as a pressure measurement catheter and for delivery of contrast media during angiographic studies.

Published July 9, 2014 · Recall initiated: May 27, 2014

Summary

FDA records medical device Class I recall Z-1981-2014, published July 9, 2014 and initiated May 27, 2014. Recalling firm: Vascular Solutions, Inc.; Status: Terminated; Reported quantity: 8,580 total (US 5714); Lot: Lot Numbers: 569436, 569437, 569962, , 569963, 569964, 569965, 570177, 570178, 570337, 570338, 570339, 570340, 570341, 570504, 570505, 570506, 570507, 570592, 570593, 570674, 570675, 570744, 570745, 570746, 570836, 570926, 570927, 570928, 570929, 571099, 571100, 571101, 571102, 571103, 571350, 571412, 571413, 571414, 571415, and 571727. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Vascular Solutions, Inc.
Distributor
Worldwide Distribution - USA nationwide including DC and Puerto Rico, AUSTRIA, CANADA, GERMANY, ISRAEL, ITALY, NETHERLANDS, SPAIN, SWEDEN SWITZERLAND, and UNITED KINGDOM.
Lot
Lot Numbers: 569436, 569437, 569962, , 569963, 569964, 569965, 570177, 570178, 570337, 570338, 570339, 570340, 570341, 570504, 570505, 570506, 570507, 570592, 570593, 570674, 570675, 570744, 570745, 570746, 570836, 570926, 570927, 570928, 570929, 571099, 571100, 571101, 571102, 571103, 571350, 571412, 571413, 571414, 571415, and 571727.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Langston Dual Lumen Catheters, 6Fr, REF 5550 and 5540. The Vascular Solutions Langston Dual Lumen Pressure Monitoring catheter is used as a pressure measurement catheter and for delivery of contrast media during angiographic studies. recalled?

Yes. FDA published recall Z-1981-2014 on July 9, 2014. Reported hazard: The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter