Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
Published May 13, 2026 · Recall initiated: March 19, 2026
Summary
FDA records medical device Class I recall Z-1981-2026, published May 13, 2026 and initiated March 19, 2026. Recalling firm: American Contract Systems Inc; Status: Ongoing; Reported quantity: 7,311 kits; Lot: ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Contract Systems Inc
- Distributor
- US distribution to MN. No OUS distribution.
- Lot
- ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays CARDIACOctober 7, 2026
- FDAMedical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACKOctober 7, 2026
- FDAMedical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ABDOMINALOctober 7, 2026
- FDAAmerican Contract System Convenience Kit, D & C PACK, Model/Catalog Number: LLPD90HSeptember 2, 2026
- FDAHalyard VA DETROIT CATH LAB PACK kit. Model Numbers: VMCD029.May 13, 2026
- FDAMedical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/CaseMay 13, 2026
Questions about this recall
Was Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ recalled?
Yes. FDA published recall Z-1981-2026 on May 13, 2026. Reported hazard: Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail