Eon Mini Neurostimulation (IPG) System (Model 3788); The product is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Published July 18, 2012 · Recall initiated: December 19, 2011
Summary
FDA records medical device Class II recall Z-1983-2012, published July 18, 2012 and initiated December 19, 2011. Recalling firm: Advanced Neuromodulation Systems Inc.; Status: Terminated; Reported quantity: 25, 255 devices; Lot: Model Number 3788. All lots manufactured prior to September 20, 2010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm has received 112 complaints of the Eon Mini IPGs that lost the ability to communicate or recharge due to a workmanship issue resulting in loss of pain relief and subsequent explant. The firm has determined the cause of these reports to be related to process variances in the positioning of the internal battery and printed circuit board, causing a short, and therefore, prematurely depletin
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Advanced Neuromodulation Systems Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Puerto Rico, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Lot
- Model Number 3788. All lots manufactured prior to September 20, 2010.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAdvanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788March 6, 2013
- FDAThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allowsOctober 17, 2012
- FDAThe Eon Mini Neurostimulation (IPG) System (Model 3788) is a 16-channel, rechargeable, multi-program systemSeptember 5, 2012
- FDAEon (Model 3716) IPG is a dual channel; 8-contact receiver designed to be connected to 4, 8, or 16 electrodeSeptember 5, 2012
- FDAStryker Orthopaedics, Definition PM Femoral Hip Stem Catalog Numbers 6265-3-099 through 6265-3-107. HowmedicaJuly 18, 2012
- FDAPudenz Flushing Valve with Integral Connectors, catalog number NL850-1330. Integra Neurosciences, Plainsboro,July 18, 2012
Questions about this recall
Was Eon Mini Neurostimulation (IPG) System (Model 3788); The product is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain. recalled?
Yes. FDA published recall Z-1983-2012 on July 18, 2012. Reported hazard: The firm has received 112 complaints of the Eon Mini IPGs that lost the ability to communicate or recharge due to a workmanship issue resulting in loss of pain relief and subsequent explant. The firm has determined the cause of these reports to be related to process variances in the positioning of the internal battery and printed circuit board, causing a short, and therefore, prematurely depletin
What should I do if I have this product?
Voluntary: Firm initiated · Letter