Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

Published July 9, 2014 · Recall initiated: May 9, 2014

Summary

FDA records medical device Class II recall Z-1985-2014, published July 9, 2014 and initiated May 9, 2014. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 10; Lot: model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.
Lot
model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department. recalled?

Yes. FDA published recall Z-1985-2014 on July 9, 2014. Reported hazard: There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.

What should I do if I have this product?

Voluntary: Firm initiated · Letter