Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.
Published July 9, 2014 · Recall initiated: May 9, 2014
Summary
FDA records medical device Class II recall Z-1985-2014, published July 9, 2014 and initiated May 9, 2014. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 10; Lot: model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.
- Lot
- model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDASynthes B37 Replacement Screws. M3.5 Screw for insertion handle guide/aiming arm, part numbers 3.113.025 andJuly 9, 2014
- FDAMicroline Surgical ReNew Laproscopic Fenestrated Grasper Forceps Tips Ref 3221July 9, 2014
Questions about this recall
Was Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department. recalled?
Yes. FDA published recall Z-1985-2014 on July 9, 2014. Reported hazard: There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.
What should I do if I have this product?
Voluntary: Firm initiated · Letter