Infusomat(R) Space Pump (US Version) with software version 686G030103 (commonly termed 'G03'), 686E030003, 686F030005, 686F030007, 686G030002, and 686G030102. Intended for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.
Published July 25, 2012 · Recall initiated: March 23, 2012
Summary
FDA records medical device Class I recall Z-1992-2012, published July 25, 2012 and initiated March 23, 2012. Recalling firm: B. Braun Medical, Inc.; Status: Terminated; Reported quantity: 38,225 pumps; Lot: Pump Model Numbers: 8713050U, 8713050U-99, 8713050US, 8713060U, and 8713060U-99 -- various serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the clip catch is broken and the door opened, free flow protection is still ensured. However, after the set is manually removed from the pump without the closure of t
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B. Braun Medical, Inc.
- Distributor
- Nationwide Distribution-including Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.
- Lot
- Pump Model Numbers: 8713050U, 8713050U-99, 8713050US, 8713060U, and 8713060U-99 -- various serial numbers.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock,September 25, 2024
- FDAInfusomat UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV-Used with an electrically-powered infusion pump forJuly 24, 2024
- FDAInfusomat UNIV. 15 DROP PUMP SET W/ 0.2 FILTER- Used with an electrically-powered infusion pump forJuly 24, 2024
- FDAInfusomat SPACE PUMP IV SET, 127 IN.-Used with an electrically-powered infusion pump for intravenous (IV)July 24, 2024
- FDAProbitotic Dietary Supplement. : iFlora® Kids Multi-Probiotic®July 25, 2012
- FDAGE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to takeJuly 25, 2012
Questions about this recall
Was Infusomat(R) Space Pump (US Version) with software version 686G030103 (commonly termed 'G03'), 686E030003, 686F030005, 686F030007, 686G030002, and 686G030102. Intended for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration. recalled?
Yes. FDA published recall Z-1992-2012 on July 25, 2012. Reported hazard: Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the clip catch is broken and the door opened, free flow protection is still ensured. However, after the set is manually removed from the pump without the closure of t
What should I do if I have this product?
Voluntary: Firm initiated · Letter