FDAFDA · Food and drugs·Recall number Z-2014-2020
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90123, UDI # 10888857108363, Size 24x6mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
Published May 27, 2020 · Recall initiated: April 13, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- K2M, Inc
- Distributor
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- Lot
- Lot # FKXC
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States