Back to search
FDAFDA · Food and drugs·Recall number Z-2021-2012

System 1 Base, 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Published July 25, 2012 · Recall initiated: June 21, 2012

Summary

FDA records medical device Class II recall Z-2021-2012, published July 25, 2012 and initiated June 21, 2012. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 648; Lot: Catalog Number: 801764 and Serial numbers: 0006-0066, 01 00-0311, 0313-0322,1001-1027,1100-1434,1436, 1440, and 1441. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Terumo® Advanced Perfusion System 1 has displayed a Red X error message on multiple module icons indicating a loss or fluctuation of internal power on one side of the system's base. Terumo Cardiovascular Systems' initial investigation has determined that there may be more than one cause for the malfunction and is investigating further to determine the root cause. Potential risks with the loss or

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV. and the countries of ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
Lot
Catalog Number: 801764 and Serial numbers: 0006-0066, 01 00-0311, 0313-0322,1001-1027,1100-1434,1436, 1440, and 1441.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was System 1 Base, 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. recalled?

Yes. FDA published recall Z-2021-2012 on July 25, 2012. Reported hazard: Terumo® Advanced Perfusion System 1 has displayed a Red X error message on multiple module icons indicating a loss or fluctuation of internal power on one side of the system's base. Terumo Cardiovascular Systems' initial investigation has determined that there may be more than one cause for the malfunction and is investigating further to determine the root cause. Potential risks with the loss or

What should I do if I have this product?

Voluntary: Firm initiated · Letter