FDAFDA · Food and drugs·Recall number Z-2032-2019

Artis zee floor MN, Model Number 10094142

Published July 31, 2019 · Recall initiated: June 12, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution
Lot
Serial Numbers: 140212 140115 140101 140205 140213 140204 140209 140219 140224 140218 140221 140220 140206 140223 140105 140226 140108 140102 140109 140100 140202 140214 140114
Risk classification
Class II
Area (FDA)
device
Country
United States