Artis zee floor MN, Model Number 10094142
Published July 31, 2019 · Recall initiated: June 12, 2019
Summary
FDA records medical device Class II recall Z-2032-2019, published July 31, 2019 and initiated June 12, 2019. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 23; Lot: Serial Numbers: 140212 140115 140101 140205 140213 140204 140209 140219 140224 140218 140221 140220 140206 140223 140105 140226 140108 140102 140109 140100 140202 140214 140114. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Nationwide Distribution
- Lot
- Serial Numbers: 140212 140115 140101 140205 140213 140204 140209 140219 140224 140218 140221 140220 140206 140223 140105 140226 140108 140102 140109 140100 140202 140214 140114
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAAXIOM Artis dTC, Model Number 7413078July 31, 2019
- FDAMonoject Standard Hypodermic Needle Item code: 8881200029July 31, 2019
Questions about this recall
Was Artis zee floor MN, Model Number 10094142 recalled?
Yes. FDA published recall Z-2032-2019 on July 31, 2019. Reported hazard: A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter