Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
Published May 10, 2017 · Recall initiated: March 23, 2017
Summary
FDA records medical device Class II recall Z-2034-2017, published May 10, 2017 and initiated March 23, 2017. Recalling firm: Sysmex America Inc; Status: Terminated; Reported quantity: 3,682 units; Lot: **Model: XN-10, Product Number: AP795756, Serial Numbers: 11001 29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sysmex America Inc
- Distributor
- Nationwide, Canada
- Lot
- **Model: XN-10, Product Number: AP795756, Serial Numbers: 11001 29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIC-Green (indocyanine green for injection, USP), 25 mg, packaged in a kit containing six vials of 25 mgMay 17, 2017
- FDADestination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices intoMay 10, 2017
- FDAAQUIOS CL Flow Cytometer, Ref no. B30166, Software Versions 2.0 and 2.0.1 Product Usage: The AQUIOS CL FlowMay 10, 2017
- FDA5.0cm Short AttachmentMay 10, 2017
- FDASMALL ATTACHMENT, J-LATCHMay 10, 2017
- FDApocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1November 6, 2013
Questions about this recall
Was Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan recalled?
Yes. FDA published recall Z-2034-2017 on May 10, 2017. Reported hazard: Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
What should I do if I have this product?
Voluntary: Firm initiated · Letter