FDAFDA · Food and drugs·Recall number Z-2034-2017

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan

Published May 10, 2017 · Recall initiated: March 23, 2017

Summary

FDA records medical device Class II recall Z-2034-2017, published May 10, 2017 and initiated March 23, 2017. Recalling firm: Sysmex America Inc; Status: Terminated; Reported quantity: 3,682 units; Lot: **Model: XN-10, Product Number: AP795756, Serial Numbers: 11001  29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Sysmex America Inc
Distributor
Nationwide, Canada
Lot
**Model: XN-10, Product Number: AP795756, Serial Numbers: 11001  29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan recalled?

Yes. FDA published recall Z-2034-2017 on May 10, 2017. Reported hazard: Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.

What should I do if I have this product?

Voluntary: Firm initiated · Letter