Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Published July 25, 2012 · Recall initiated: May 4, 2012
Summary
FDA records medical device Class II recall Z-2037-2012, published July 25, 2012 and initiated May 4, 2012. Recalling firm: Hitachi Medical Systems America Inc; Status: Terminated; Reported quantity: Six (6) units; Lot: Serial #'s S5001 - S5011. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hitachi Medical Systems America Inc
- Distributor
- US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.
- Lot
- Serial #'s S5001 - S5011
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHitachi Echelon MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI System - C-Spine CoilFebruary 1, 2017
- FDAHitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended toSeptember 10, 2014
- FDAHudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to addJuly 25, 2012
- FDAAlere Triage Tox+MTD Drug Screen, PN 94400 Product Usage: The Alere Triage TOX Drug Screen is a fluorescenceJuly 25, 2012
Questions about this recall
Was Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering. recalled?
Yes. FDA published recall Z-2037-2012 on July 25, 2012. Reported hazard: Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit