FDAFDA · Food and drugs·Recall number Z-2037-2014

GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476. A diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Published July 16, 2014 · Recall initiated: June 18, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Distributed in the states of CA, CO, IL, OK, OR and WI.
Lot
GEMINI TF PET/CT16: 882470 7019,7039,7099,7588,7593,7598,7599,7600,7604,7605,7596M GEMINI TF PET/CT 64: 882471 7027,7044,7146,7153,7178,7182,7188,7578,7579,7585,7587,7589,7592,7594,7595,7602, 7603, 7606 GEMINI TF Big Bore: 882476 9019,9227,9232,9233
Risk classification
Class II
Area (FDA)
device
Country
United States