GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476. A diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Published July 16, 2014 · Recall initiated: June 18, 2014
Summary
FDA records medical device Class II recall Z-2037-2014, published July 16, 2014 and initiated June 18, 2014. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 33; Lot: GEMINI TF PET/CT16: 882470 7019,7039,7099,7588,7593,7598,7599,7600,7604,7605,7596M GEMINI TF PET/CT 64: 882471 7027,7044,7146,7153,7178,7182,7188,7578,7579,7585,7587,7589,7592,7594,7595,7602, 7603, 7606 GEMINI TF Big Bore: 882476 9019,9227,9232,9233. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Distributed in the states of CA, CO, IL, OK, OR and WI.
- Lot
- GEMINI TF PET/CT16: 882470 7019,7039,7099,7588,7593,7598,7599,7600,7604,7605,7596M GEMINI TF PET/CT 64: 882471 7027,7044,7146,7153,7178,7182,7188,7578,7579,7585,7587,7589,7592,7594,7595,7602, 7603, 7606 GEMINI TF Big Bore: 882476 9019,9227,9232,9233
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAReliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and UtensilJuly 16, 2014
- FDAPALL Medical Breathing Circuit Filter, REF BB50T and BB50T-BULK , Rx ONLY Single use, bi-directionalJuly 16, 2014
Questions about this recall
Was GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476. A diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. recalled?
Yes. FDA published recall Z-2037-2014 on July 16, 2014. Reported hazard: When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.
What should I do if I have this product?
Voluntary: Firm initiated · Letter