FDAFDA · Food and drugs·Recall number Z-2039-2012
Alere Triage Tox+MTD Drug Screen, PN 94400 Product Usage: The Alere Triage TOX Drug Screen is a fluorescence immunoassay to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses. This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/ Mass Spectrometry/ Mass Spectrometry (LC/MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods. A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.
Published July 25, 2012 · Recall initiated: May 14, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Alere San Diego, Inc.
- Distributor
- US Nationwide Distribution
- Lot
- Lot/Unit Codes: W50216B, W50222B, W50245B, W50250B, W50809B, W50810B, W50812B, W50813B, W50816B, W50817B, W50818B, W50819B, W50820B, W50821B, W50822B, W50823B, W50826B, W50829B, W50830B, W50831B, W50832B, W50837B, W50838B, W50842B, W50843B, W50844B, W50846B, W50848B, W50851B, W50862B, W50866B, W50869B, W50870B, W50871B, W50872B, W50873B.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States