MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)
Published June 19, 2024 · Recall initiated: May 10, 2024
Summary
FDA records medical device Class II recall Z-2041-2024, published June 19, 2024 and initiated May 10, 2024. Recalling firm: Canon Medical System, USA, INC.; Status: Ongoing; Reported quantity: 153; Lot: Model/UDI-DI/Serial Numbers: MRT-1550 (MEXL-1550)/04987670103360/POA21X2002, S5A1982001, S5B22Z2041, S5B2172018, S2C2252024, S5A20Z2009, S2C21Y2021, S2B2052010, S5B2272031, POA2262005, S5B2292036, S5B2272029, S5B2282034, S5B22Y2040, S5B2292035, S6B2252008, S5A20Z2010, POA2252004, POA2182001, S1C2152071, S5B2352047, S2C2252023, S5B2122013, S6B2262010, S2C2192020, S6B2292013, S5B2132014, S6B2322018, S5B2292038, S6B2312016, S2B2052011, S2C2162018, S4A1992013, S3A19Z2013, S3A1982009, S5B2182021, S6B2322017, S5A2092008, S6B2362019, S5B22Z2042, S6B2222005, S2B2042009, S2B20Z2015, S5B2162017, S6B2262011, S5B2272030, S6B22Z2014, S5B2282033, S8B2272009, S5A2062007, S2B2072012, S2B2072013, S6B2222006, S6B2272012, S2B2072014, S6A20Z2001, POA2272006, S5A19Z2003, S1C2142070, POA2212003, S2C2142017, S5B2362048, S2A1962005, S6B2182004, S5B2292037, S6B2252007, S5B2282032, S5B2142015, S5B2382050, S2A18Z2003, S6B2112002, S6B2382021, S1B2042051, S8B2212007, S1B2062057, S1A1922022, S2A18Z2001, S1B2092067, S1C2242080, S5B2172019, S6B2252009, S6B2372020, S2A18Z2002, S2A1942004, S2C2212022, S2C2272025, S6B2392022, S2C2162019, S2C2132016, S4B2062020, S1C2262081, PDA21Y2001, S6B22Z2015, S5B22Z2043, S5B2242027, MRT-2020 (MEXL-1520)/04987670101632/S1F18X2580, S1B1492068, S1D17Y2461, S1F2052753, S1C1722390, S1C1682336, S1E1842517, S1E1842527, S1C1672334, MRT-1504 (MEXL-1504)/04987670102400/U5B15Z2001, MRT-3010 (MEXL-3010)/04987670101649/A5D14X2032, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5F1562034, A5B11Y2005, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5D1362021, A5C1252012, A5C1292015, A5D13X2024, A5C1422029, A5B11Z2007, A5C1212008, MRT-3020 (MEXL-3020)/04987670102615/D5A2082005, D5A2092007, D5B2152013, S5B17X2034, D7B2182017, 5DB2152002, S5B18X2046, S5A17X2033, D5A2062003, S5A1742021, D5B2212019, D5B2192016, S5A1662001, S5C18Y2048, S5A1732019, S5A1732018, S5A17Y2035, S7B1892004, S5B1842043, D7B2152015, S5C1972054, 5DB2172003. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Canon Medical System, USA, INC.
- Distributor
- US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR
- Lot
- Model/UDI-DI/Serial Numbers: MRT-1550 (MEXL-1550)/04987670103360/POA21X2002, S5A1982001, S5B22Z2041, S5B2172018, S2C2252024, S5A20Z2009, S2C21Y2021, S2B2052010, S5B2272031, POA2262005, S5B2292036, S5B2272029, S5B2282034, S5B22Y2040, S5B2292035, S6B2252008, S5A20Z2010, POA2252004, POA2182001, S1C2152071, S5B2352047, S2C2252023, S5B2122013, S6B2262010, S2C2192020, S6B2292013, S5B2132014, S6B2322018, S5B2292038, S6B2312016, S2B2052011, S2C2162018, S4A1992013, S3A19Z2013, S3A1982009, S5B2182021, S6B2322017, S5A2092008, S6B2362019, S5B22Z2042, S6B2222005, S2B2042009, S2B20Z2015, S5B2162017, S6B2262011, S5B2272030, S6B22Z2014, S5B2282033, S8B2272009, S5A2062007, S2B2072012, S2B2072013, S6B2222006, S6B2272012, S2B2072014, S6A20Z2001, POA2272006, S5A19Z2003, S1C2142070, POA2212003, S2C2142017, S5B2362048, S2A1962005, S6B2182004, S5B2292037, S6B2252007, S5B2282032, S5B2142015, S5B2382050, S2A18Z2003, S6B2112002, S6B2382021, S1B2042051, S8B2212007, S1B2062057, S1A1922022, S2A18Z2001, S1B2092067, S1C2242080, S5B2172019, S6B2252009, S6B2372020, S2A18Z2002, S2A1942004, S2C2212022, S2C2272025, S6B2392022, S2C2162019, S2C2132016, S4B2062020, S1C2262081, PDA21Y2001, S6B22Z2015, S5B22Z2043, S5B2242027, MRT-2020 (MEXL-1520)/04987670101632/S1F18X2580, S1B1492068, S1D17Y2461, S1F2052753, S1C1722390, S1C1682336, S1E1842517, S1E1842527, S1C1672334, MRT-1504 (MEXL-1504)/04987670102400/U5B15Z2001, MRT-3010 (MEXL-3010)/04987670101649/A5D14X2032, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5F1562034, A5B11Y2005, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5D1362021, A5C1252012, A5C1292015, A5D13X2024, A5C1422029, A5B11Z2007, A5C1212008, MRT-3020 (MEXL-3020)/04987670102615/D5A2082005, D5A2092007, D5B2152013, S5B17X2034, D7B2182017, 5DB2152002, S5B18X2046, S5A17X2033, D5A2062003, S5A1742021, D5B2212019, D5B2192016, S5A1662001, S5C18Y2048, S5A1732019, S5A1732018, S5A17Y2035, S7B1892004, S5B1842043, D7B2152015, S5C1972054, 5DB2172003
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000CNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000FNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000VNovember 4, 2026
- FDAAlphenix INFX-8000C/B5October 14, 2026
- FDAMyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicatedJune 19, 2024
- FDAOMTech Desktop Laser EngraverJune 19, 2024
Questions about this recall
Was MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020) recalled?
Yes. FDA published recall Z-2041-2024 on June 19, 2024. Reported hazard: The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
What should I do if I have this product?
Voluntary: Firm initiated · Letter