Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.
Published July 15, 2015 · Recall initiated: September 23, 2013
Summary
FDA records medical device Class II recall Z-2043-2015, published July 15, 2015 and initiated September 23, 2013. Recalling firm: Volcano Corp; Status: Terminated; Reported quantity: 227 total units affected by the issue - all models; Lot: s5 Part number 804200-002 and 804200-014; s5x Part number 807300-001; CORE MOBILE Part Number 400-0100.01. All software versions. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Volcano Corp
- Distributor
- Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.
- Lot
- s5 Part number 804200-002 and 804200-014; s5x Part number 807300-001; CORE MOBILE Part Number 400-0100.01. All software versions
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v,50/60hZ, 1000 VAOctober 7, 2026
- FDAArthroscopy Pack - contains Devon Light Glove Used during surgeryJuly 15, 2015
- FDAA.L.P.S. Foot Locking Calcaneus Plate,Small-Right. Provides the orthopaedic surgeon a means of bone fixationJuly 15, 2015
- FDABlepharoplasty Pack - contains Devon Light Glove Used during surgeryJuly 15, 2015
- FDAPowerPICC SOLO Catheter with Tip Location System (TLS) Stylet 4F Nurse Full Tray with Microintroducer (withJuly 15, 2015
- FDAShoulder Pack - contains Devon Light Glove Used during surgeryJuly 15, 2015
Questions about this recall
Was Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures. recalled?
Yes. FDA published recall Z-2043-2015 on July 15, 2015. Reported hazard: A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
What should I do if I have this product?
Voluntary: Firm initiated · Letter