ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Published August 1, 2012 · Recall initiated: June 25, 2012
Summary
FDA records medical device Class II recall Z-2073-2012, published August 1, 2012 and initiated June 25, 2012. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 18; Lot: ARTISTE, Part #: 08139789, ONCOR Avant Garde, Part #: 05863472, ONCOR Impression Plus, Part #: 05857912, ONCOR Expression, Part #: 07360717. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Worldwide Distribution, including the state of Florida (USA) and the countries of Germany, India, Poland, Thailand, and the Republic of Korea.
- Lot
- ARTISTE, Part #: 08139789, ONCOR Avant Garde, Part #: 05863472, ONCOR Impression Plus, Part #: 05857912, ONCOR Expression, Part #: 07360717
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPRIMUS, ONCOR, ARTISTE Linac systems with component 550 TxT Treatment Table. TT-A, TT-D, TT-M or TT-S TableJanuary 29, 2014
- FDASIMVIEW NT and SIMVIEW NT with IDI: October 01, 2012 Siemens ceased Simview distribution. Manufactured by:January 29, 2014
- FDASIEMENS brand ARTISTE Linear Accelerator Linac systems with component: any photon unflat beam option (MultipleDecember 11, 2013
- FDASIEMENS brand ONCOR series Linear Accelerator Linac systems with component: any photon unflat beamDecember 11, 2013
- FDAProduct is labeled in part: "***OPTETRAK RBK***TIBIAL TRAY***FINNED, CEMENTED***CoCr Alloy***SIZE 2F/1T***REFAugust 1, 2012
- FDABridgePoint Medical, Stingray Catheter, REF M - 1000, Sterilized with Ethylene Oxide Gas, Prescription Only,August 1, 2012
Questions about this recall
Was ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. recalled?
Yes. FDA published recall Z-2073-2012 on August 1, 2012. Reported hazard: Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter