Back to search
FDAFDA · Food and drugs·Recall number Z-2073-2012

ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Published August 1, 2012 · Recall initiated: June 25, 2012

Summary

FDA records medical device Class II recall Z-2073-2012, published August 1, 2012 and initiated June 25, 2012. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 18; Lot: ARTISTE, Part #: 08139789, ONCOR Avant Garde, Part #: 05863472, ONCOR Impression Plus, Part #: 05857912, ONCOR Expression, Part #: 07360717. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Worldwide Distribution, including the state of Florida (USA) and the countries of Germany, India, Poland, Thailand, and the Republic of Korea.
Lot
ARTISTE, Part #: 08139789, ONCOR Avant Garde, Part #: 05863472, ONCOR Impression Plus, Part #: 05857912, ONCOR Expression, Part #: 07360717
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. recalled?

Yes. FDA published recall Z-2073-2012 on August 1, 2012. Reported hazard: Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.

What should I do if I have this product?

Voluntary: Firm initiated · Letter