ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585
Published August 14, 2019 · Recall initiated: June 20, 2019
Summary
FDA records medical device Class I recall Z-2074-2019, published August 14, 2019 and initiated June 20, 2019. Recalling firm: St Jude Medical Inc.; Status: Terminated; Reported quantity: 15; Lot: Serial Number: 9836058, 9836059, 9836061, 9836062, 9836063, 9836064, 9836065, 9836066, 9836067, 9836219, 9836220, 9836221, 9838312, 9838544, 9838546. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
What to do
Voluntary: Firm initiated · Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- St Jude Medical Inc.
- Distributor
- Worldwide distribution. US nationwide, and countries: DE, FR, IT, ES, NL, PT, SE, HU, DK, PL, IL, GP, DZ, CH, SA, SK, LU, NZ, VN, and NO
- Lot
- Serial Number: 9836058, 9836059, 9836061, 9836062, 9836063, 9836064, 9836065, 9836066, 9836067, 9836219, 9836220, 9836221, 9838312, 9838544, 9838546
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAELLIPSE VR, REF: CD1377-36QC Implantable cardioverter defibrillators (ICDs)August 14, 2019
- FDAELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs)August 14, 2019
- FDAYXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testingAugust 14, 2019
- FDAYXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testingAugust 14, 2019
- FDAMerlin PCS programmer (Model # 3650), Software (Model # 3330)July 4, 2018
- FDAUnify, Sterile EO, Model #/ Part #: CD3231-40/100029240, 100030104, 100041904; CD3231-40Q/100029281,July 4, 2018
Questions about this recall
Was ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585 recalled?
Yes. FDA published recall Z-2074-2019 on August 14, 2019. Reported hazard: Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
What should I do if I have this product?
Voluntary: Firm initiated · Visit