PrepStain PM Kit, T-AG (Catalog 490588); PrepStain PM Kit Europe (Catalog 490671); PrepStain Prep Only PM Kit, T-AG (Catalog 490954); PrepStain PM Kit XLP, T-US-II (Catalog 490424); PrepStain Prep PM Kit XLP T-US-II (Catalog 491118); PrepStain Prep Only PM Kit XLP, T-US-II (490423); PrepStain Prep Only PM Kit, T-US (Catalog 490429); PrepStain PM Kit, T-US (Catalog 490626); PrepStain PM Kit Europe, T-US (Catalog 490428). The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in the screening and detection of cervical cancer, pre-cancerous lesions, atypical cells and all other cytologic categories as defined by The Bethesda System for Reporting Cervical/Vaginal Cytology Diagnoses. The 1.2 DiTi is the tip on the PrepStain to which a pipette tip attaches and dispenses patient samples onto a glass microscope slide.
Published August 1, 2012 · Recall initiated: December 8, 2011
Summary
FDA records medical device Class II recall Z-2088-2012, published August 1, 2012 and initiated December 8, 2011. Recalling firm: Tripath Imaging, Inc.; Status: Terminated; Reported quantity: 1166 units; Lot: SAP Part numbers: 490423, 490424, 490428, 490588, 490626, 490954 and 490671. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Tripath Imaging, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the states AL, AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY, and WV and the countries of Belgium, Brazil, Canada, China, Columbia, Hong Kong, India, Japan, Korea, Peru, Singapore, Taiwan and Victoria (Australia).
- Lot
- SAP Part numbers: 490423, 490424, 490428, 490588, 490626, 490954 and 490671.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPrepStain Kit: 1.2 DiTi Cone. The PrepStain System is a liquid-based thin layer cell preparation process. TheAugust 1, 2012
- FDAPrepStain Upgrade Kit: 1.3.0.3/1.2 DiTi Cone. The PrepStain System is a liquid-based thin layer cellAugust 1, 2012
- FDAPrepStain Slide Processor. The PrepStain System is a liquid-based thin layer cell preparation process. TheAugust 1, 2012
- FDAABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland CarrigtwohillAugust 1, 2012
- FDARejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, IrelandAugust 1, 2012
- FDAPrepStain Installation Kit (PreCoat), Catalog/Model # 490528June 27, 2012
Questions about this recall
Was PrepStain PM Kit, T-AG (Catalog 490588); PrepStain PM Kit Europe (Catalog 490671); PrepStain Prep Only PM Kit, T-AG (Catalog 490954); PrepStain PM Kit XLP, T-US-II (Catalog 490424); PrepStain Prep PM Kit XLP T-US-II (Catalog 491118); PrepStain Prep Only PM Kit XLP, T-US-II (490423); PrepStain Prep Only PM Kit, T-US (Catalog 490429); PrepStain PM Kit, T-US (Catalog 490626); PrepStain PM Kit Europe, T-US (Catalog 490428). The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in the screening and detection of cervical cancer, pre-cancerous lesions, atypical cells and all other cytologic categories as defined by The Bethesda System for Reporting Cervical/Vaginal Cytology Diagnoses. The 1.2 DiTi is the tip on the PrepStain to which a pipette tip attaches and dispenses patient samples onto a glass microscope slide. recalled?
Yes. FDA published recall Z-2088-2012 on August 1, 2012. Reported hazard: Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions.
What should I do if I have this product?
Voluntary: Firm initiated · Letter