Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter.

Published July 30, 2014 · Recall initiated: May 10, 2012

Summary

FDA records medical device Class II recall Z-2089-2014, published July 30, 2014 and initiated May 10, 2012. Recalling firm: Perouse Medical; Status: Terminated; Reported quantity: 2,496 units; Lot: Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Perouse Medical
Distributor
US.
Lot
Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448.
Risk classification
Class II
Area (FDA)
device
Country
France

Related recalls

Questions about this recall

Was Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter. recalled?

Yes. FDA published recall Z-2089-2014 on July 30, 2014. Reported hazard: Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail