Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.
Published July 29, 2015 · Recall initiated: June 10, 2015
Summary
FDA records medical device Class II recall Z-2097-2015, published July 29, 2015 and initiated June 10, 2015. Recalling firm: Maquet Cardiovascular Us Sales, Llc; Status: Terminated; Reported quantity: 12 units; Lot: SN 00048. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Distributor
- Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.
- Lot
- SN 00048
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - ProductDecember 23, 2020
- FDAHLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 -December 23, 2020
- FDAYuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before,October 28, 2020
- FDAMaquet/Getinge Rotaflow System Console US (115 V / 60 Hz) - Product Usage: Instructions for Use - CentrifugalSeptember 23, 2020
- FDAMetoprolol Succinate Extended-Release Tablets, USP, 50 mg Tablets, Packaged in A) 100 Count Bottles (NDC:July 29, 2015
- FDAStryker SelectCore Variable Throw Biopsy Device, 14G, 10CM Product Usage: The Stryker SelectCore VariableJuly 29, 2015
Questions about this recall
Was Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co. recalled?
Yes. FDA published recall Z-2097-2015 on July 29, 2015. Reported hazard: There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
What should I do if I have this product?
Voluntary: Firm initiated · Letter