Aesculap SIBD 4 degree Trial Implant (SJ768T-US) and Aesculap SIBD 9 degree Trial Implant (SJ780T-US) Product Usage: The Aesculap Implant Systems SIBD Trial Implant instruments assist the surgeon in selecting the correct size implant for the patient

Published August 8, 2012 · Recall initiated: July 12, 2012

Summary

FDA records medical device Class II recall Z-2114-2012, published August 8, 2012 and initiated July 12, 2012. Recalling firm: Aesculap Implant Systems LLC; Status: Terminated; Reported quantity: 5; Lot: 4 degree - SJ768T-US and 9 degree SJ780T-US. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The 4 degree and 9 degree trial instruments (SJ768T-US/SJ780T-US) are correctly dimensioned, but the part number etched on the instrument may be incorrect. The 4 degree trial instrument (SJ768T-US) may be marked with the incorrect part number of the 9 degree trail instrument (SJ780T-US) and the 9 degree trial instrument (SJ780T-US) may be marked with the incorrect part number of the 4 degree tria

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aesculap Implant Systems LLC
Distributor
US Nationwide Distribution - including the states of CO, OH, OK and TX.
Lot
4 degree - SJ768T-US and 9 degree SJ780T-US
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Aesculap SIBD 4 degree Trial Implant (SJ768T-US) and Aesculap SIBD 9 degree Trial Implant (SJ780T-US) Product Usage: The Aesculap Implant Systems SIBD Trial Implant instruments assist the surgeon in selecting the correct size implant for the patient recalled?

Yes. FDA published recall Z-2114-2012 on August 8, 2012. Reported hazard: The 4 degree and 9 degree trial instruments (SJ768T-US/SJ780T-US) are correctly dimensioned, but the part number etched on the instrument may be incorrect. The 4 degree trial instrument (SJ768T-US) may be marked with the incorrect part number of the 9 degree trail instrument (SJ780T-US) and the 9 degree trial instrument (SJ780T-US) may be marked with the incorrect part number of the 4 degree tria

What should I do if I have this product?

Voluntary: Firm initiated · Letter