FDAFDA · Food and drugs·Recall number Z-2120-2012
Octane-C Cervical, 6¿ Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7).
Published August 8, 2012 · Recall initiated: June 15, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Exactech, Inc.
- Distributor
- Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo.
- Lot
- Sizes: 5 mm Catalog # 05-010-02-0605 All serial numbers 6 mm Catalog # 05-010-02-0606 All serial numbers 7 mm Catalog # 05-010-02-0607 All serial numbers 8 mm Catalog # 05-010-02-0608 All serial numbers 9 mm Catalog # 05-010-02-0609 All serial numbers 10mm Catalog # 05-010-02-0610 All serial numbers 11mm Catalog # 05-010-02-0611 All serial numbers 12mm Catalog # 05-010-02-0612 All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States