FDAFDA · Food and drugs·Recall number Z-2120-2012

Octane-C Cervical, 6® Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7).

Published August 8, 2012 · Recall initiated: June 15, 2012

Summary

FDA records medical device Class II recall Z-2120-2012, published August 8, 2012 and initiated June 15, 2012. Recalling firm: Exactech, Inc.; Status: Terminated; Reported quantity: 54,877 units for all products in Recall Event; Lot: Sizes: 5 mm Catalog # 05-010-02-0605 All serial numbers 6 mm Catalog # 05-010-02-0606 All serial numbers 7 mm Catalog # 05-010-02-0607 All serial numbers 8 mm Catalog # 05-010-02-0608 All serial numbers 9 mm Catalog # 05-010-02-0609 All serial numbers 10mm Catalog # 05-010-02-0610 All serial numbers 11mm Catalog # 05-010-02-0611 All serial numbers 12mm Catalog # 05-010-02-0612 All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Exactech, Inc.
Distributor
Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo.
Lot
Sizes: 5 mm Catalog # 05-010-02-0605 All serial numbers 6 mm Catalog # 05-010-02-0606 All serial numbers 7 mm Catalog # 05-010-02-0607 All serial numbers 8 mm Catalog # 05-010-02-0608 All serial numbers 9 mm Catalog # 05-010-02-0609 All serial numbers 10mm Catalog # 05-010-02-0610 All serial numbers 11mm Catalog # 05-010-02-0611 All serial numbers 12mm Catalog # 05-010-02-0612 All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Octane-C Cervical, 6® Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7). recalled?

Yes. FDA published recall Z-2120-2012 on August 8, 2012. Reported hazard: Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year

What should I do if I have this product?

Voluntary: Firm initiated · Letter