L-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) Left (003-E-00001L); b) Right (003-E-00001R)

Published June 13, 2018 · Recall initiated: April 6, 2018

Summary

FDA records medical device Class II recall Z-2130-2018, published June 13, 2018 and initiated April 6, 2018. Recalling firm: Orthofix, Inc; Status: Terminated; Reported quantity: 30; Lot: (UDI)/Lot: a) (18056099647830) B1108952; b) (18056099647847) B1108955. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Orthofix, Inc
Distributor
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Lot
(UDI)/Lot: a) (18056099647830) B1108952; b) (18056099647847) B1108955
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was L-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) Left (003-E-00001L); b) Right (003-E-00001R) recalled?

Yes. FDA published recall Z-2130-2018 on June 13, 2018. Reported hazard: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

What should I do if I have this product?

Voluntary: Firm initiated · Letter