cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.

Published June 20, 2018 · Recall initiated: April 3, 2018

Summary

FDA records medical device Class II recall Z-2136-2018, published June 20, 2018 and initiated April 3, 2018. Recalling firm: Roche Diagnostics Corporation; Status: Terminated; Reported quantity: 35; Lot: Serial numbers: 1744-08 1744-09 1744-10 1758-08 1606-09 1752-05 1752-06 1630-08 1738-01 1744-07 1758-09 1759-04 1758-06 1758-07 1743-09 1743-10 1752-04 1759-05 17C5-06 17C5-07 1615-08 1612-10 17C5-09 1744-05 1744-06 1743-07 1743-08 1744-03 1744-04 1759-06 17C5-05 17C5-08 17C5-10 17C6-01 17C6-02. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diagnostics Corporation
Distributor
US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.
Lot
Serial numbers: 1744-08 1744-09 1744-10 1758-08 1606-09 1752-05 1752-06 1630-08 1738-01 1744-07 1758-09 1759-04 1758-06 1758-07 1743-09 1743-10 1752-04 1759-05 17C5-06 17C5-07 1615-08 1612-10 17C5-09 1744-05 1744-06 1743-07 1743-08 1744-03 1744-04 1759-06 17C5-05 17C5-08 17C5-10 17C6-01 17C6-02
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement. recalled?

Yes. FDA published recall Z-2136-2018 on June 20, 2018. Reported hazard: The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.

What should I do if I have this product?

Voluntary: Firm initiated · Letter