KODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Canada) --- A permanently installed diagnostic x-ray system composed of 3 main components: an operator console, U-Arm assembly with x-ray tube including a collimator, and a stationary generator.
Published August 13, 2014 · Recall initiated: May 30, 2014
Summary
FDA records medical device Class II recall Z-2137-2014, published August 13, 2014 and initiated May 30, 2014. Recalling firm: Carestream Health Inc.; Status: Terminated; Reported quantity: Domestic: 100 units; Foreign: 46 units; Lot: Service Code 1534. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A control system failure may cause the U-Arm to move downward unexpectedly when the Radiology Technologist is positioning the equipment using the "Up" or "Down" control buttons. Downward movement occurs at three times the normal rate of speed and will continue until the device control button is released, the Emergency Stop button is activated or the equipment Bump Sensor contacts an object.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Carestream Health Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Canada, China, Italy, Germany, Greece, Kuwait, Spain, South Africa, and United Arab Emirates.
- Lot
- Service Code 1534
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAUltra-speed Dental Film, Catalog Number/REF 834 8658, KODAK and Carestream DENTAL labels --- Carestream HealthAugust 13, 2014
- FDAINSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream HealthAugust 13, 2014
- FDASerrated Soft Tissue Shields. Orthopedic surgical instruments, part of the Silex Sacroiliac Fusion System. TheAugust 13, 2014
- FDAAdvanDx Quick FISH Fixation Kit for BC ( Blood Cultures), 50 tests. In vitro diagnostic. The QuickFISHAugust 13, 2014
- FDAModel Kodak 2200 Intraoral X-Ray System, Catalog Numbers 5303177, 5303169, 5303151, XR-5154323 -- CarestreamMarch 12, 2014
- FDAModel Kodak 2100 Intraoral X-Ray System, Catalog Numbers 5303144, 8011579 -- Carestream Health, Inc. 150March 12, 2014
Questions about this recall
Was KODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Canada) --- A permanently installed diagnostic x-ray system composed of 3 main components: an operator console, U-Arm assembly with x-ray tube including a collimator, and a stationary generator. recalled?
Yes. FDA published recall Z-2137-2014 on August 13, 2014. Reported hazard: A control system failure may cause the U-Arm to move downward unexpectedly when the Radiology Technologist is positioning the equipment using the "Up" or "Down" control buttons. Downward movement occurs at three times the normal rate of speed and will continue until the device control button is released, the Emergency Stop button is activated or the equipment Bump Sensor contacts an object.
What should I do if I have this product?
Voluntary: Firm initiated · Letter