TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used with the Tablo Hemodialysis System.
Published June 26, 2024 · Recall initiated: April 30, 2024
Summary
FDA records medical device Class II recall Z-2150-2024, published June 26, 2024 and initiated April 30, 2024. Recalling firm: Outset Medical, Inc.; Status: Ongoing; Reported quantity: 431; Lot: UDI-DI: 00850001011303. All PN-0006813 serial numbers with power cords: PN-0005746, PN-0006578, PN-0005747. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Outset Medical, Inc.
- Distributor
- US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK, MA, WV, IN, KY, AZ, LA, NE, KS, NJ, CT, NM, CO, SC, TX, PA, NY, OH.
- Lot
- UDI-DI: 00850001011303. All PN-0006813 serial numbers with power cords: PN-0005746, PN-0006578, PN-0005747
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo HemodialysisOctober 9, 2024
- FDAMONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximationJune 26, 2024
- FDACentrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient supportJune 26, 2024
- FDAMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligationJune 26, 2024
- FDAHRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900June 26, 2024
- FDATablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001April 24, 2024
Questions about this recall
Was TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used with the Tablo Hemodialysis System. recalled?
Yes. FDA published recall Z-2150-2024 on June 26, 2024. Reported hazard: A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.
What should I do if I have this product?
Voluntary: Firm initiated · Letter