FDAFDA · Food and drugs·Recall number Z-2167-2013
Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.
Published September 18, 2013 · Recall initiated: April 17, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A complaint investigation has found that product fractures can occur at the distal end of the catheter under load.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Xlumena, Inc.
- Distributor
- Worldwide Distribution - USA including AZ, CA, FL, GA, IN, NY,SC, WA, PA, IL, CO, and internationally to Netherlands, Germany, Spain, Denmark, and Switzerland.
- Lot
- All lots are involved are being recalled, including FGS00048, FGS0005, FGS00058, FGS00059, FGS00069, FGS00073, FGS00081, FGS00086, FGS00091
- Risk classification
- Class II
- Area (FDA)
- device
- Country