Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.
Published September 18, 2013 · Recall initiated: April 17, 2013
Summary
FDA records medical device Class II recall Z-2167-2013, published September 18, 2013 and initiated April 17, 2013. Recalling firm: Xlumena, Inc.; Status: Terminated; Reported quantity: 148 units; Lot: All lots are involved are being recalled, including FGS00048, FGS0005, FGS00058, FGS00059, FGS00069, FGS00073, FGS00081, FGS00086, FGS00091. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A complaint investigation has found that product fractures can occur at the distal end of the catheter under load.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Xlumena, Inc.
- Distributor
- Worldwide Distribution - USA including AZ, CA, FL, GA, IN, NY,SC, WA, PA, IL, CO, and internationally to Netherlands, Germany, Spain, Denmark, and Switzerland.
- Lot
- All lots are involved are being recalled, including FGS00048, FGS0005, FGS00058, FGS00059, FGS00069, FGS00073, FGS00081, FGS00086, FGS00091
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOutlook Safety Infusion System Burette Set. For intravenous fluid administration.September 18, 2013
- FDAOPERON D850 Surgical Table Product Usage: Surgical TableSeptember 18, 2013
- FDABaxa Vial Adapter, 14mm - Luer Lock, Order No. REF: 90, Quantity: 50 each, Single Use. Product Code H93890.September 18, 2013
- FDAADVIA Centaur Folate (100 Tests), SMN 10310308, Cat No. 06367974 In vitro diagnostic use in the quantitativeSeptember 18, 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 18, 2013
- FDAADVIA Centaur Folate (500 Tests) SMN 10325366, Cat No. 06891541 In vitro diagnostic use in the quantitativeSeptember 18, 2013
Questions about this recall
Was Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures. recalled?
Yes. FDA published recall Z-2167-2013 on September 18, 2013. Reported hazard: A complaint investigation has found that product fractures can occur at the distal end of the catheter under load.
What should I do if I have this product?
Voluntary: Firm initiated · Letter