Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option.
Published August 20, 2014 · Recall initiated: May 28, 2014
Summary
FDA records medical device Class II recall Z-2174-2014, published August 20, 2014 and initiated May 28, 2014. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 160; Lot: material numbers: 01940035, 04504200, 05672977, 05693908, 05857912, 05857920, 08139789, 08319758, 08490005, 08515520, 09401654, 09401746, 09822693, with serial numbers below 3094. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Nationwide and ROW, including Canada and Mexico.
- Lot
- material numbers: 01940035, 04504200, 05672977, 05693908, 05857912, 05857920, 08139789, 08319758, 08490005, 08515520, 09401654, 09401746, 09822693, with serial numbers below 3094.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAMedtronic MiniMed Implantable Insulin Pump, Model No. MMT-2007DAugust 20, 2014
- FDADX-D 600 - DXD Imaging Package Product Usage:Is indicated for use in providing diagnostic quality images toAugust 20, 2014
Questions about this recall
Was Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option. recalled?
Yes. FDA published recall Z-2174-2014 on August 20, 2014. Reported hazard: A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter