Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option.

Published August 20, 2014 · Recall initiated: May 28, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Nationwide and ROW, including Canada and Mexico.
Lot
material numbers: 01940035, 04504200, 05672977, 05693908, 05857912, 05857920, 08139789, 08319758, 08490005, 08515520, 09401654, 09401746, 09822693, with serial numbers below 3094.
Risk classification
Class II