Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diagnostics Dimension(R) clinical chemistry analyzers are floor model, fully automated, microprocessor-controlled, integrated instrument systems that use prepackaged Flex(R) reagent test cartridges to measure a variety of analytes in human body fluids.
Published August 15, 2012 · Recall initiated: July 17, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to overheat and potentially create a fire hazard. RMS without a power cord retention clip may be affected.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Distributed in the US including Washington, DC, Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WI, WV, and WY and the Virgin and Pacific Islands.
- Lot
- Serial numbers 99060001 - 99062554
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States