Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA),subcutaneous, epidural, enteral, or irrigation of fluid spaces. The Pump module is indicated for use on adults, pediatrics, and neonates.
Published August 22, 2012 · Recall initiated: June 29, 2012
Summary
FDA records medical device Class I recall Z-2185-2012, published August 22, 2012 and initiated June 29, 2012. Recalling firm: CareFusion 303, Inc.; Status: Terminated; Reported quantity: 102,248 units; Lot: All serial numbers associated with the Alaris Pump Module Model 8100: 8100 DEMO PUMP MODULE; 8100 Global Pump Module; 8100 IOC Pump Module V8.5.29.0; 8100 LVP Training Unit; 8100 Pump Module V8.5.29.0; 8100 Pump Module V8.5.29.0 (V9); Medley Patient Care System LVP; Model 8100 Trade-In Material; RCND 8100 Pump Module V8.5.29.0. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of: South Africa and Canada. Additional Countries: Netherlands.
- Lot
- All serial numbers associated with the Alaris Pump Module Model 8100: 8100 DEMO PUMP MODULE; 8100 Global Pump Module; 8100 IOC Pump Module V8.5.29.0; 8100 LVP Training Unit; 8100 Pump Module V8.5.29.0; 8100 Pump Module V8.5.29.0 (V9); Medley Patient Care System LVP; Model 8100 Trade-In Material; RCND 8100 Pump Module V8.5.29.0
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinetJune 10, 2026
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinetJune 10, 2026
- FDABD Pyxis Mini Main REF: 349 Medication cabinetJune 10, 2026
- FDABD Pyxis MedStation ES REF: 323 Medication cabinetJune 10, 2026
- FDAIMPAX Cardiovascular (CV) Reporting Reporting tool used for structured reporting and structured report outputsAugust 22, 2012
- FDATAUT, Cholangiogram Catheter Intraducer, Peritoneal Catheter, Teleflex Medical, Research Triangle Park, NC. AnAugust 22, 2012
Questions about this recall
Was Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA),subcutaneous, epidural, enteral, or irrigation of fluid spaces. The Pump module is indicated for use on adults, pediatrics, and neonates. recalled?
Yes. FDA published recall Z-2185-2012 on August 22, 2012. Reported hazard: The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.
What should I do if I have this product?
Voluntary: Firm initiated · Letter